Pickering Laboratories carbamate columns are guaranteed to produce the separation of carbamate residues, specified by EPA 531.1, EPA 531.2 and AOAC 985.23 methods.

Post-column conditions for pesticide analysis:

Reagent 1: Hydrolysis reagent CB130 or CB130.2 Reagent 2: 100 mg of OPA, 2 g Thiofluor™ in 950 mL of CB910 Reactor 1: 100 °C, 0.5 mL Reactor 2: ambient. 0.1 mL Reagents flow rate: 0.3 mL/min

Detection:

Fluorometer: λex 330 nm, λem 465 nm

The recommended gradient conditions are subject to change without notice. This may happen because of lot specific changes in the columns, or improvements in the overall method.

The recommended gradient for the column will always be included in the column package, and it supersedes the information in this catalog.

[tabby title=”High Resolution”]

High Resolution/Capacity Carbamate Column 0846250 (4.6 X 250 mm), C8, 5 µm

chart-ccg-1846250b

1  Aldicarb sulfoxide 2  Aldicarb sulfone 3  Oxamyl 4  Methomyl 5  3-Hydroxycarbofuran 6  Aldicarb 7  Propoxur 8  Carbofuran   9  Carbaryl 10  1-Naphthol 11  Methiocarb 12  BDMC (internal standard)

CONDITIONS FOR AQUEOUS SAMPLES

Time (Min) WATER % MeOH %
0 100 0
1 100 0
1.1 82 18
36 30 70
39 30 70
39.1 0 100
41 0 100
41.1 100 0
55 100 0

FLOW RATE: 1 mL/min, COLUMN TEMP.: 42 ºC, INJECTION VOLUME: Up to 400 µL

CONDITIONS FOR METHANOLIC SAMPLES

Time (Min) WATER % MeOH %
0 85 15
1 85 15
44 25 75
44.1 0 100
49 0 100
49.1 85 15
57 85 15

FLOW RATE: 1 mL/min, COLUMN TEMP.: 42 ºC, INJECTION VOLUME: 10 µL

[tabby title=”Rapid Analysis”]

Rapid Analysis Carbamate Column 1846150 (4.6 X 150 mm), 5 µm

chart-ccg-1846150a

CONDITIONS FOR AQUEOUS SAMPLES

Time (Min) WATER % MeOH %
0 100 0
1 100 0
1.1 88 12
26 30 70
26.1 0 100
28 0 100
28.1 100 0
38 100 0

FLOW RATE: 1 mL/min, COLUMN TEMP.: 42 ºC, INJECTION VOLUME: Up to 250 µL

chart-ccg-1846150b
1  Aldicarb sulfoxide 2  Aldicarb sulfone 3  Oxamyl 4  Methomyl 5  3-Hydroxycarbofuran 6  Aldicarb 7  Propoxur 8  Carbofuran   9  Carbaryl 10  1-Naphthol 11  Methiocarb 12  BDMC (internal standard)

CONDITIONS FOR METHANOLIC SAMPLES

Time (Min) WATER % MeOH %
0 85 15
0.5 85 15
28.5 30 70
28.6 0 100
33.5 0 100
33.6 85 15
41 85 15

FLOW RATE: 1 mL/min, COLUMN TEMP.: 42 ºC, INJECTION VOLUME: 10 µL

[tabby title=”Expanded Resolution”]

Expanded Resolution Carbamate Column 0840250 (4.0 X 250 mm), 5 µm

chart-ccg-0840250a

CONDITIONS FOR WATER/METHANOL GRADIENT

Time (Min) WATER % MeOH %
0 85 15
2 85 15
42 30 70
46 30 70
46.1 0 100
51 0 100
51.1 85 15
59 85 15

FLOW RATE: 0.8 mL/min, COLUMN TEMP.: 50 ºC, INJECTION VOLUME: 10 µL for methanolic and up to 400 µL for aqueous samples

chart-ccg-0840250b
1  Aldicarb sulfoxide 2  Aldicarb sulfone 3  Oxamyl 4  Methomyl 5  3-Hydroxycarbofuran 6  Aldicarb 7  Propoxur 8  Carbofuran   9  Carbaryl 10  1-Naphthol 11  Methiocarb 12  BDMC (internal standard)

CONDITIONS FOR WATER / ACETONITRILE GRADIENT

Time (Min) WATER % ACN %
0 90 10
2 90 10
46 49 51
46.1 30 70
49 30 70
49.1 90 10
59 90 10

FLOW RATE: 0.8 mL/min, COLUMN TEMP.: 50 ºC, INJECTION VOLUME: 10 µL for methanolic and up to 400 µL for aqueous samples

[tabby title=”23 Carbamates”]

Separation of 23 Carbamates Using Column 0840250 (4.0 X 250 mm)

chart-ccg-0840250c
1 Aldicarb sulfoxide 2 Aldicarb sulfone 3 Oxamyl 4 Methomyl 5 3-Hydroxy 6 Aldicarb 7 Propoxur 8 Carbofuran    9 Carbaryl 10 Methiocarb 11 BDMC internal standard 12 Butocarboxim sulfoxide 13 Butocarboxim sulfone 14 Ethiofencarb sulfoxide 15 Ethiofencarb sulfone 16 Butocarboxim 17 Ethiofencarb 18 Thiofanox sulfoxide 19 Thiofanox sulfone 20 Methiocarb sulfoxide 21 Methiocarb sulfone 22 3-Ketocarbofuran 23 Thiofanox N 1-Naphthol

CONDITIONS FOR WATER/METHANOL GRADIENT

Time (Min) WATER % MeOH %
0 85 15
2 85 15
42 30 70
46 30 70
46.1 0 100
51 0 100
51.1 85 15
59 85 15

FLOW RATE: 0.8 mL/min, COLUMN TEMP.: 50 ºC, INJECTION VOLUME: 10 µL for methanolic and up to 400 µL for aqueous samples